Charged Plate Monitors
Supplier: Trek Inc. (Advanced Energy)
Technology type: Charged Plate Monitors
TREK’s charged plate monitors are made to evaluate and monitor the performance of air ionisers. This technology delivers static charge neutralisation and tests the efficiency of an ioniser's ion production.
Operating principle:
Charged plate monitors (CPM’s) evaluate the ability of air ionisers to neutralise static charge. A CPM charges an isolated, conductive plate that is charged to a known level. Then the air ions produced by the ioniser neutralises the charge on the isolated plate, and the time required to discharge the isolated plate is measured.
Models 156, 157, and 158A feature three primary modes of operation: a negative (-) discharge mode, a positive (+) discharge mode, and a balanced mode. In addition, 158a offers an ALL-function mode. The ALL-function mode permits automated tests for +DECAY, -DECAY, and BALANCE which are then saved to memory.
Range features
A high level overview of what this range offers
- Monitored voltage range: 0 to ±1100 V DC or peak AC
- Bandwidth (-3dB): DC to 80Hz
- Decay mode thresholds: Programmable from 1 to ±1000V in 1V increments
- Programmable user-friendly compact design with superior data collection and storage capability
- Programmable start and stop voltage for (+) and (-) decay time ioniser tests
- Calibration traceable to NIST
What's in this range?
All the variants in the range and a comparison of what they offer
Model |
Monitored voltage range |
Large signal bandwidth |
Decay mode threshold |
---|---|---|---|
PD04002A |
0 to ±55VDC or peak AC |
DC to greater than 200Hz (-3dB) |
- |
156A |
0 to ±1100VDC or peak AC |
DC to 10Hz (measured at 2000 V p-p) |
Programmable from 1 to ±1000V in 1V increments (accurate within ±1V of programmed voltage |
157 |
0 to ±1020VDC or peak AC |
DC to 80Hz (-3dB) |
Programmable from 1 to ±1000V in 1V increments (accurate within ±1V of programmed voltage) |
158A |
0 to ±1100VDC or peak AC |
DC to 80Hz (-3dB) |
Programmable from 1 to ±1000V in 1V increments |
Monroe 288C |
0 to ±1200VDC or peak AC |
DC to 10Hz (measured at 2000 V p-p) |
- |